Examples of using Linearity verification in English and their translations into Russian
{-}
-
Official
-
Colloquial
FTIR: linearity verification.
Measurement systems that require linearity verification.
A linearity verification shall be performed for each measurement system listed in table 7.
Measurement systems that require linearity verifications.
Introduction A linearity verification shall be performed for each measurement system listed in Table 7.
Table 8.2.- Measurement systems that require linearity verifications.
The linearity verification shall be run for each normally used operating range with the following steps.
Table 8.2 indicates measurement systems that require linearity verifications.
The linearity verification shall be run for each normally used operating range with the following steps.
Min" refers to the minimum reference value used during the linearity verification;
For the other instruments and systems, the linearity verification shall be done within 370 days before testing.
For the linearity verification, a NH3 gas that meets the specifications of paragraph A.7.4.2.7. shall be used.
Max" generally refers to the maximum reference value used during the linearity verification.
For conducting the linearity verification according to paragraph 9.2.1., the gas divider shall be accurate to within 1 per cent.
Each flow meter used in a particulate sampling anddouble dilution system shall be subjected to a linearity verification as required by the instrument manufacturer.
A linearity verification shall consist of introducing a series of at least 10 reference values to a measurement system, unless otherwise specified.
For temperature measurement systems with thermocouple, RTD, or thermistor sensors,the calibration of the system shall be performed as described in paragraph 8.1.4.4. for linearity verification.
A linearity verification shall be performed more frequently if the instrument manufacturer recommends it or based on good engineering judgment;
For temperature measurement systems with sensors like thermocouples, RTDs,and thermistors, the linearity verification may be performed by removing the sensor from the system and using a simulator in its place.
A linearity verification of the analyzer shall be performed using propane in air and purified synthetic air according to paragraph 9.2.1.3.
At the outlet of the gas-division system,a gas analyzer is recommended to be connected that meets the linearity verification described in this section and has not been linearized with the gas divider being verified.
The linearity verification shall be repeated after correcting the deficiency to ensure that the measurement system meets the linearity criteria.
These need to satisfy other criteria such as in paragraphs 8.1.8.6.(periodic calibration) and 8.2.1.2.(validation) for varying dilution PFD, andparagraph 8.1.4.5. as well as Table 8.2(linearity verification) and paragraph 8.1.8.5.7.(verification) for constant dilution PFD.
The linearity verification shall be performed at least every 12 months or whenever a system repair or change is made that could influence the calibration.
For the other instruments and systems, the linearity verification shall be done as required by internal audit procedures, by the instrument manufacturer or in accordance with ISO 9000 requirements.
Linearity verification shall be done as required by internal audit procedures, by the instrument manufacturer or in accordance with ISO 9000 requirements.
The linearity verification in accordance with paragraph 9.2.1. of this annex, shall be performed at least every 12 months or whenever a system repair or change is made that could influence calibration.
The linearity verification in accordance with paragraph 9.2.1. shall be performed for the gas analyzers within 35 days before testing or whenever a system repair or change is made that could influence calibration.
The linearity verification according to paragraph 9.2.1. shall be performed for the gas analysers at least every 3 months or whenever a system repair or change is made that could influence calibration.
A linearity verification shall be performed on each measurement system listed in table 8.2 at least as frequently as indicated in the table, consistent with measurement system manufacturer recommendations and good engineering judgment.