NH3 calibration gas.
U Dynamic viscosity of calibration gas kg/m·s.
Calibration gas port as option.
Probe and calibration gas control.
Calibration gas- propane; propane bypasses NMC.
The flowing the NO2 calibration gas shall be stopped;
Calibration gas- methane; methane passes through NMC.
Reference data of each calibration gas used;
Calibration gas 2 CLOSED CLOSED 2.9 Sample stream connection.
Reference data of each calibration gas used.
The calibration gas supply pressure will be in the range 1-5 psig.
A dilution device can be applied for diluting a 100 per cent calibration gas to required concentration.
The calibration gas consisting of propane/air bypasses the NMC;
The NOx analyser shall be in the NO mode so that the calibration gas does not pass through the converter.
The calibration gas consisting of methane/air passes through the NMC.
NOx and purified nitrogen(the amount of NO2 contained in this calibration gas must not exceed 5% of the NO content);
Used as a calibration gas and to balance gas calibration mixtures.
NO in nitrogen(the amount of NO2 contained in this calibration gas shall not exceed 5 per cent of the NO content);
Calibration gas" means a gas mixture used to calibrate gas analysers.
Real and indicated value of each calibration gas used together with the percentage differences;
The test shall be considered acceptable if the difference between the pre-test andpost-test results is less than 2 per cent of the calibration gas value.
Regardless of the calibration gas composition, it shall be calibrated on a carbon number basis of one C1.
Vii It shall be immediately switched back to overflowing the NO2 calibration gas used to establish xNOxref.
Methane calibration gas must be flown through the FID with and without bypassing the NMC and the two concentrations recorded.
NO and purified nitrogen(the amount of NO2 contained in this calibration gas shall not exceed 5 per cent of the NO content);
Ethane efficiency Ethane calibration gas must be flown through the FID with and without bypassing the NMC and the two concentrations recorded.
NO and purified nitrogen.The amount of NO2 contained in this calibration gas shall not exceed 5 per cent of the NO content.
The methane/air calibration gas shall be flowed to the FID through the NMC and bypassing the NMC and the two concentrations recorded.
The analyser shall be calibrated(zeroed/ calibrated) in the NO2 mode using the NO2 certified concentration in the calibration gas cylinder.
Ii NO2 calibration gas(balance gas of dry air) that has an NO2 concentration that is near the maximum expected during testing shall be selected.